Dear customer, when you use the CleanTrust Touch Luminometer, this software can help you realize the data transmission between the device and the desktop computer, and complete the data analysis.

Tietojen synkronointi

Toiminnon esittely
Analyzing software data synchronization, the synchronization operation will synchronize the historical data, detection points, inspectors, detection plans, and threshold data in the instrument to the software. It can also synchronize the detection points, inspectors, detection plans, and thresholds in the analysis software to the instrument. The initial interface is shown in Figure 1Scigiene-PC-Version-Analysis-Software-fig (1)

Tietojen synkronointi
There are two data synchronization methods: WIFI connection and USB connection. The two data synchronization modes cannot be used simultaneously.

WIFI-yhteys

WIFI connection method push data:

  1. Click on the “Analysis Software Synchronization” icon to enter the analysis software synchronization interface, as shown in Figure 2. Fill in the software service IP and service port number displayed in the data synchronization interface of the computer analysis software into the corresponding box of the instrument analysis software synchronization interface;
  2. On the data synchronization interface of the computer analysis software, click “Start Reception”;
  3. Click the “Push Data” button on the instrument side;
  4. After synchronization is completed, the synchronized data can be viewed on the computer analysis software.

WIFI connection method to pull data:

  1. Click the “Pull Data” button on the device side to enter the data pull interface, as shown in Figure 3, and then click the “Pull Data” button at the bottom of the interface to open the service;
  2. Click the “Start Push” button in the analysis software, enter the device IP, and click the “Push Data” button to start the data transmission service;
  3. After the data transmission is completed, the pulled data can be viewed laitteeseen.

USB-tila
USB connection method push data:

  1. Please connect the instrument to a desktop computer with installed analysis software using a Type-C USB cable;
  2. Click on the “Analysis Software Synchronization” icon on the device to enter the analysis software synchronization interface, and then click on “No network push”;
  3. Select “ATP3500” in the “Equipment” column of the computer analysis software, and click “Netless Reception” to complete the synchronization operation of instrument data to the analysis software data.

USB connection method for pulling data:

  1. Connect the device and the desktop computer where the analysis software is located using a Type-C USB data cable;
  2. On the device side, click the “Pull Data” button to enter the pull data interface. On the computer side, select “ATP3500” in the “Device” column of the analysis software, click the “No network push” button, and the data will start to be transmitted. After the transmission is completed, there will be relevant prompt messages;
  3. Click the “No Network Pull” button on the device side to complete a network free data push.Scigiene-PC-Version-Analysis-Software-fig (2)Scigiene-PC-Version-Analysis-Software-fig (4)

Detection Result

Toiminnon esittely
The detection result in the analysis software is the management of the synchronized data, including data query, query condition clearing, data printing, data export and deletion. The interface is shown in Figure 4 as below:Scigiene-PC-Version-Analysis-Software-fig (5)

The query conditions include detection time, detection point, inspector, inspection plan, retest, inspection result, whether to delete, device name, and device code. You can select multiple search conditions except whether to delete or retest. You need to select a specific date when checking the time selection.

Kysely
Select the query conditions, click the query button to query.

Selkeä
Click the Clear button to clear the selected query.

Painaa
Click the Print button, and the “Select Data” window will pop up. Select the name of the data column to be printed, and click the OK button. The interface is shown in Figure 5 below:Scigiene-PC-Version-Analysis-Software-fig (6)

Viedä
Click the Export button, select the save path, and export the data in the table in excel form.

Poistaa
Drag the scroll bar to find “Delete” and click Delete to delete the data.

Tietojen tilastot

Toiminnon esittely
The data statistics function in the analysis software of t h e CleanTrust Touch Luminometer is the statistics of all the synchronized data. It contains statistics by year, quarter, month, week, inspection point, inspector, test plan, test result, and equipment name.Annual, quarterly, monthly, and weekly statistics are displayed in line charts and pie charts. Detection point, inspector, inspection plan, inspection result, and device name are shown in bar charts and pie charts (The interface is shown in Figure 6 below).Scigiene-PC-Version-Analysis-Software-fig (7)

The query conditions include detection time, detection point, inspector, inspection plan, retest, inspection result, whether to delete, device name, and device code. You can select multiple search conditions except whether to delete or retest. You need to select a specific date when checking the time selection.

Kysely
Select the query conditions, click the query button to query.

Selkeä
Click the Clear button to clear the selected query.

Painaa
Click the Print button to print the table together with the statistics shown.

Chart Switching
Click “Pie chart”, “Line chart” or “Bar chart” to switch the statistical chart display. As shown in Figure 7, 8 and 9 below.Scigiene-PC-Version-Analysis-Software-fig (8)Scigiene-PC-Version-Analysis-Software-fig (9)

Laitehallinta

Toiminnon esittely
Analysis software devices management is the form of previously synchronized connected devices. As shown in Figure 10 below.Scigiene-PC-Version-Analysis-Software-fig (10)

Kysely
Enter the device code, click query to query.

Poistaa
Napsauta Poista poistaaksesi laitteen.

Käyttäjien hallinta

Toiminnon esittely
Analysis software user management, is a management mechanism of user permission and data security. The permission data management mechanism is not enabled by default.

Enabling mechanism: Add an administrator account and enable the login mode (As shown in Figure 11 below).Scigiene-PC-Version-Analysis-Software-fig (11)

Permission differentiation:

  1. Administrator: can add ordinary users, management of ordinary user permissions.
  2. Ordinary user: Do not have user management permissions. When the administrator assigns deletion permissions to ordinary users, they have the permissionto delete data; otherwise, they do not have the permission to delete data.

Kysely
Enter the username and click query to query.

Käyttäjän lisäys
Click the Add button, and the user Add window will pop up. If the user management data is empty, the first user is an administrator by default, and only one administrator can exist. Click OK to complete the addition. As shown in FIG. 12 and FIG. 13 below.Scigiene-PC-Version-Analysis-Software-fig (12)Scigiene-PC-Version-Analysis-Software-fig (3)

User Modification
This operation can be performed only after the administrator logs in. The administrator password is required for editing the administrator. Click User modification to pop up the user editing window, enter the modification information and click OK to modify successfully. The following is the picture of each editing interface:Scigiene-PC-Version-Analysis-Software-fig (13)Scigiene-PC-Version-Analysis-Software-fig (14) Scigiene-PC-Version-Analysis-Software-fig (15)

Poistaa
Administrators cannot be deleted, but common users can. Click Delete to delete theuser.

Kirjaudu sisään
On the user management page, select Enable login and restart the software. As shown in Figure 17 below:Scigiene-PC-Version-Analysis-Software-fig (16)

Enter the username and password and click Login to enter the software according to the role of the current user. The login page is shown in Figure 18.Scigiene-PC-Version-Analysis-Software-fig (17)

esimample, a common user has the Delete permission and can delete data in detectionresults and device management. If a device is assigned, the data is viewed based on the owned device.

Muista salasana
If Remember Password is selected, you do not need to enter the username andpassword for the next login (As shown in Figure 19).Scigiene-PC-Version-Analysis-Software-fig (18)

Salasana unohtunut
If you forget the password, the administrator must perform operations. Click Forgotpassword, pop-up answer the question window, the question is set when registeringthe administrator, the administrator enters the answer. As shown in Figure 20:Scigiene-PC-Version-Analysis-Software-fig (19)

Click the OK button. The Change Password window is displayed. Change the userpassword and click OK to complete the change successfully. As shown in Figure 21:Scigiene-PC-Version-Analysis-Software-fig (20)

Toiminnon esittely
The detection points in the analysis software are for the management of the synchronized data and the newly added detection points in the analysis software. Itincludes detection point query, query condition clearing, detection point addition,detection point editing, and deletion. The interface is shown in Figure 22:Scigiene-PC-Version-Analysis-Software-fig (21)

Kysely
Enter the name of the detection point and click the “Query” button.

Lisätä
Click “Add” button, enter the detection point information, and click “OK” button to complete the addition.

Muokata
Click “Edit” and enter the modified detection point information. If the detection point is used in the detection plan, the detection point name cannot be modified. Click the “OK” button to complete the modification.

Poistaa
Click “Delete” to delete the detection point. If the detection point is used in the detection plan, it cannot be deleted.

Selkeä
Click the “Clear” button to clear the check point query condition

Tarkastaja

Toiminnon esittely
The inspector in the analysis software is the management of the synchronizeddata and the new inspector in the analysis software. It includes inspector query, query condition clearing, inspector add, inspector edit, and delete, as shown in Figure 23:Scigiene-PC-Version-Analysis-Software-fig (22)

Kysely
Enter the name of the inspector and click the “Query” button.

Lisätä
Click “Add” button, enter the inspector information, and click “OK” button to complete the addition.

Muokata
Click “Edit”, enter the modified inspector information, and click “OK” button tocomplete the editing.

Poistaa
Click “Delete” to delete the inspector.

Selkeä
Click the “Clear” button to clear the inspector query condition.

Inspection Plan

Toiminnon esittely
The detection plan in analysis software is the management of the synchronizeddata and the new detection plan in the analysis software. It includes query of test plan, clearing of query conditions, adding test plan, editing test plan, and deleting test plan.

The interface is shown in Figure 24:Scigiene-PC-Version-Analysis-Software-fig (23)

Kysely
Enter the test plan name and click the “Query” button.

Lisätä
Click “Add” button, enter the test plan information, and click “OK” button tocomplete the addition.

Muokata
Click “Edit”, enter the modified detection plan information, and click “OK”button to complete the editing.

Poistaa
Click “Delete” to delete the detection plan.

Selkeä
Click the “Clear” button to clear the query conditions of the detection plan.

Kynnys

Toiminnon esittely
The threshold in the analysis software is the management of the synchronized data and the new threshold in the analysis software. It includes inspector query, query condition clearing, inspector add, inspector edit, and delete, as shown in Figure 25:Scigiene-PC-Version-Analysis-Software-fig (24)

Kysely
Enter a threshold name and click the Query button.

Lisätä
Click the “Add” button, enter the threshold information, and click the “OK” button to complete the addition.

Muokata
Click “Edit” and enter the modified threshold information. If the threshold is used in the detection point, the threshold name cannot be modified. Click “OK” tocomplete the editing.

Poistaa
Click Delete to delete the threshold. If the threshold is used in the detection point,it cannot be deleted.

Selkeä
Click the Clear button to clear the threshold query conditions.

Asetukset

Chinese-English shift
Click on the upper right corner of the software and select Chinese and English to switch.

Versiot
Napsauta ” Scigiene-PC-Version-Analysis-Software-fig (25)” to expand the drop-down list, select the version and the version number window will.

Auttaa
napsauta "Scigiene-PC-Version-Analysis-Software-fig (25) ” to expand the drop-down and select Help to open the instruction manual directly.

Tietojen siirto
Data migration is to search for the detection result data in the old version wheninstalling a new version. Migrate the detection result data of the old version to the newversion. When installing the new version, if the detection result data of the old version is found, a prompt will be given. At this point, you can choose to migrate or not to migrate. If you do not migrate, manual migration is required.

Manual migration of data:
Napsauta ” Scigiene-PC-Version-Analysis-Software-fig (25)” to expand the drop-down. Select data migration to open the data migration interface (as shown in Figure 27). Select the upper folder of the file where the software is located (Example: C:\Program Files (x86) \Yizhi Technology). Just click the “Data Migration” button.Scigiene-PC-Version-Analysis-Software-fig (26)

Ota yhteyttä

  • 1295 Morningside Avenue, yksikkö 16-18
  • Scarborough, ON Canada
  • Puhelin: 416-261-4865 Faksi: 416-261-7879
  • www.scigiene.com

UKK

Q: Can I use WIFI and USB synchronization simultaneously?

A: No, WIFI and USB synchronization modes cannot be used simultaneously.

Q: What query conditions are available for detection results?

A: Query conditions include detection time, detection point, inspector, inspection plan, retest, inspection result, whether to delete, device name, and device code.

Asiakirjat / Resurssit

PDF thumbnailPC Version Analysis Software
Instruction Manual · 2025-S2, PC Version Analysis Software, Version Analysis Software, Analysis Software

Viitteet

Esitä kysymys

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Esitä kysymys

Ask a question about setup, compatibility, troubleshooting, or anything missing from this manual.